Product Details
G5 VENTILATOR
Patient Age Group Paediatric, Adult and Infant
Tidal Volume Adult/Pediatric: 20 to 2000 ml, Neonatal: 2 to 200 ml
Ventilation Mode (S)CM, SIMV, VS, APVcmv, APVsimv, P-CMV, P-SIMV, DuoPAP, APRV, SPONT, ASV INTELLiVENT-ASV, NIV, NIV-ST, nCPAP-PS, Hi Flow O2
Performance Data
Maximum Inspiratory Flow
180 L/min peak flow, max. 120 L/min continuous flow
Electrical
External hot-swappable battery (optional)
14.4 V DC, 6.6 Ah, Battery type: Lithium Ion, Backup Time: Backup time typically 1 hours, Recharge time: 7 hours With external charger: 3 hours
Standards and Approvals
Classification
Class IIb, continuously operating according to EC directive 93/42/EEC
Certification
EN 60601-1:2006/A1:2013, IEC 60601-1-2:2014, ANSI/AAMI ES60601-1:2005/(R)2012, ISO, 80601-2-12:2011, CAN/CSA-C22.2 NO. 60601-1:14, EN ISO 5356-1:2015, ISO 80601-2-55:2011
Declaration
The HAMILTON-G5 was developed in accordance with pertinent international standards and FDA guidelines. The ventilator is manufactured within an EN ISO 13485 and EN ISO 9001, Council Directive 93/42/EEC, Annex II, Article 1 certified quality management system. The ventilator meets the Essential Requirements of Council Directive 93/42/EEC, Annex I.
Electromagnetic Compatibility
According to IEC 60601-1-2:2014
Safety Class
Class I, Type B applied part (ventilator breathing system, VBS), type BF applied parts CO2 sensor including CO2 module connector, humidifier, Aerogen system, nebulizer, and SpO2 sensor including SpO2 adapter, continuous operation according to IEC 60601-1
Degree of Protection
IP21
Pneumatic Performance
Oxygen Pressure
2 to 6 bar / 29 to 87 psi
Oxygen Connector
DISS (CGA 1240) or NIST (optional), NF (optional)
Air Supply
Input Pressure: 2.8 to 6 bar / 41 to 87 psi, Connector: CGA 1160-A
Inspiratory Outlet (To Patient Port)
Connector: ISO ID15/OD22 Conical
Expiratory Outlet (From Patient Port)
Connector (on expiratory valve): ISO ID15/OD22 conical, Exhaust Port: OD30
Graphical Patient Data
Intelligent Panel
Dynamic Lung4, Vent Status, ASV Graph5, ASV Monitor, SMPs (Secondary monitoring parameter)
Waveform
Paw, Flow, Volume, Off, PCO21, FCO21, Plethysmogram2, Pes (Paux)3, Ptranspulm
Trend
1, 3, 12, 24, or 96-h trend data for a selected parameter or combination of parameters
Loops
Paw/Volume, Paw/Flow, Volume/Flow, Volume/PCO2
Additional Details
Power Supply
100 to 240 V AC ±10%, 50/60 Hz, 210 VA maximum
Display Size
15 inch XGA
Display Type
Color TFT, LCD Touchscreen
Battery Backup
Typical 1 hours, Recharge time: 15 hours, Battery Type: Lead-acid
Weight
38 kg (83.8 lb), 57 kg (125.6 lb) with standard trolley, monitor, ventilation unit, The standard trolley can accommodate a maximum safe working load of 80 kg (176 lb). The universal trolley can accommodate a maximum safe working load of 140 kg (308 lb)
Apnea Backup Type
On, Off
Inspiratory Pressure Limit
0 to 120 cm H2O
Alarms
High: Apnea time (s), Exp ManVel high/low (l/min), Oxygen high/low (%), Pressure high/low (cmH2O), Flow sensor calibration needed, Exhalation obstructed, Disconnection, Oxygen supply failed Medium: Total high/low (b/min), PetCO2 high/low (mmHg), Pressure limitation (cmH2O), Vt high/low (ml), SpO2 high/low, SpOC high/low,%leak, High PEEP, Loss of PEEP, Pulse high/low, Check flow sensor for water Low: High SpO2, Loss of external power, Cuff leak
I:E Ratio
Adult/Ped: 1:9 to 4:1
Automatic Expiratory Base Flow
Adult/Pediatric. Pressure trigger: 1 L/min, Flow trigger setting ≤2 L/min: 4 L/min, Flow trigger setting >2 L/min: 2 Flow trigger, Trigger OFF: 1 L/min, IntelliSync: variable Neonatal. Pressure trigger: 1 L/min, Flow trigger setting ≤1 L/min: 2 L/min, Flow trigger setting >1 L/min: 2 Flow trigger (max. 6 L/min), Trigger OFF: 1 L/min
Means of Inspiratory Triggering
Flow Pressure or Optional IntelliSync Trigger Control
Means of Expiratory Triggering
ETS or Optional IntelliSync Control
Minimum expiratory time
20% of cycle time; 0.1 to 0.8 sec
Oxygen Mixer Accuracy
± (Volume Fraction of 2.5% + 2.5% of Actual Reading)
Preoperational Checks
Tightness Test, Flow Sensor/O2 Sensor/CO2 Sensor Calibration
Heliox
Input pressure: 2.8 to 6 bar / 41 to 87 psi, Connector: CGA 1180-A (optional)
Features
HIgh Flow Oxygen Therapy
Non Invasive Ventilation
P/V Tool Pro for lung assessment and recruitment
Transpulmonary pressure measurement
Integrated Humidifier Control
Automated ventilation with INTELLiVENT-ASV
Applications
ICU, Critical Care and Transport
Product Description
The HAMILTON-G5 was designed for the most complex, critically ill patients in all ICU settings where lung protection is of paramount importance. The comprehensive monitoring package and state-of-the-art diagnostic tools for lung assessment support you in making the best possible clinical decisions for your patient. A wide range of additional options and features, such as Heliox and high flow oxygen therapy, allows you to tailor the HAMILTON-G5 to fit the requirements of your ICU and meet the needs of each individual patient.