C3 VENTILATOR
Patient Age Group Paediatric, Adult and Infant
Tidal Volume Adult/Pediatric: 20 to 2000 ml, Neonatal: 2 to 300 ml
Ventilation Mode (S)CMV, SIMV, APVcmv / (S)CMV+, APVsimv / SIMV+, PCV+, PSIMV+, DuoPAP, APRV, SPONT, ASV INTELLiVENT-ASV, NIV, NIV-ST, nCPAP-PS, HiFlowO2
Performance Data
Maximum Inspiratory Flow
240 L/min (150 L/min with 100% O2)
Parameter
PEEP/CPAP
Adult/Ped: 0 to 35 cm H2O, Neonatal: 0 to 25 cm H2O
Control
Flow Trigger
Adult/Ped: 1 to 20 L/min, Neonatal: 0.1 to 5 L/min
Flow Pattern
Square, 50% decelerating, Sine, 100% decelerating
Standards and Approvals
Classification
Class IIb, continuously operating according to EC directive 93/42/EEC
Certification
IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2014, ANSI/AAMI ES60601-1:2005/(R)2012, ISO 80601-2-12:2011, CAN/CSA-C22.2 NO. 60601-1:14, EN ISO 5356-1:2015, ISO 80601-2-55:2011
Declaration
The HAMILTON-C3 was developed in accordance with pertinent international standards and FDA guidelines. The ventilator is manufactured within an EN ISO 13485 and EN ISO 9001, Council Directive 93/42/EEC, Annex II, Article 3 certified quality management system. The ventilator meets the Essential Requirements of Council Directive 93/42/EEC, Annex I
Electromagnetic Compatibility
According to IEC 60601-1-2:2014
Safety Class
Class II, Type B applied part (ventilator breathing system, VBS), type BF applied parts CO2 sensor including CO2 module connector, humidifier, nebulizer, and SpO2 sensor including SpO2 adapter, continuous operation according to IEC 60601-1
Pneumatic Performance
Oxygen Pressure
2.8 to 6 bar / 41 to 87 psi
Oxygen Connector
DISS (CGA 1240) or NIST
Air Supply
Integrated ultra-quiet turbine
Inspiratory Outlet (To Patient Port)
Connector: ISO 15 mm ID/22 mm OD conical
Expiratory Outlet (From Patient Port)
Connector (on expiratory valve): ISO 15 mm ID/22 mm OD conical
Graphical Patient Data
Intelligent Panel
Dynamic Lung2, Vent Status, ASV Graph3
Waveform
Pressure, Flow, Volume, PCO21, FCO21, Plethysmogram1, Ptrachea
Trend
1, 6, 12, 24, or 72-h trend data for a selected parameter or combination of parameters
Loops
Pressure/Volume, Pressure/Flow, Volume/Flow, Volume/PCO21, Volume/FCO21
Additional Details
Power Supply
100 to 240 V AC, 50/60 Hz or 12 to 24 V DC
Display Size
12.1 inch (307.3 mm) diagonal
Display Type
Color TFT
Battery Backup
3.5 hours with one battery / ≥7 hours with two batteries
Weight
Ventilation unit: 9.5 kg (21 lb); with trolley: 37 kg (81.6 lb)
Alarms
High: Apnea time (s), Exp ManVel high/low (l/min), Oxygen high/low (%), Pressure high/low (cmH2O), Flow sensor calibration needed, Exhalation obstructed, Disconnection, Oxygen supply failed, Medium: Total high/low (b/min), PetCO2 high/low (mmHg), Pressure limitation (cmH2O), Vt high/low (ml), SpO2 high/low, SpOC high/low,%leak, High PEEP, Loss of PEEP, Pulse high/low, Check flow sensor for water, Low: High SpO2, Loss of external power, Cuff leak
I:E Ratio
Adult/Ped: 1:9 to 4:1, Neonatal: 1:9 to 4:1
Automatic Expiratory Base Flow
Fixed at 6 L/min
Means of Inspiratory Triggering
Flow trigger or pressure trigger control
Means of Expiratory Triggering
Flow cycle (ETS)
Minimum expiratory time
20% of cycle time; 0.2 to 0.8 s
Oxygen Mixer Accuracy
± (Volume fraction of 2.5% + 2.5% of actual reading)
Preoperational Checks
Tightness test, Flow sensor/O2 sensor/CO2 sensor calibration
Features
HIgh Flow Oxygen Therapy
High Performance Non Invasive Ventilation
P/V Tool Pro for lung assessment and recruitment
Transpulmonary pressure measurement
Integrated High Performance Turbine
Automated ventilation with INTELLiVENT-ASV
Applications
ICU, Critical Care and Transport
Product Description
The HAMILTON-C3 includes Hamilton Medical's standardized Ventilation Cockpit user interface and the unique intelligent ventilation mode, Adaptive Support Ventilation (ASV). The HAMILTON-C3S edition includes the optional INTELLiVENT-ASV mode. This unique mode takes the proven concepts of ASV one step further: INTELLiVENT-ASV automatically controls the patient's ventilation and oxygenation based on targets set by the clinician and on physiologic input from the patient. The HAMILTON-C3 includes the optional Protective Ventilation Tool (P/V Tool Pro) to assess recruitability. The P/V Tool Pro also provides an easy and repeatable method for performing recruitment maneuvers. The HAMILTON-C3 optionally provides an integrated high flow oxygen therapy mode with a flow of up to 80 l/min.