• C3 VENTILATOR HAMILTON


C3 VENTILATOR 

Patient Age Group  Paediatric, Adult and Infant

Tidal Volume  Adult/Pediatric: 20 to 2000 ml, Neonatal: 2 to 300 ml

Ventilation Mode (S)CMV, SIMV, APVcmv / (S)CMV+, APVsimv / SIMV+, PCV+, PSIMV+, DuoPAP, APRV, SPONT, ASV INTELLiVENT-ASV, NIV, NIV-ST, nCPAP-PS, HiFlowO2

Performance Data

Maximum Inspiratory Flow

240 L/min (150 L/min with 100% O2)

Parameter

PEEP/CPAP

Adult/Ped: 0 to 35 cm H2O, Neonatal: 0 to 25 cm H2O

Control

Flow Trigger

Adult/Ped: 1 to 20 L/min, Neonatal: 0.1 to 5 L/min

Flow Pattern

Square, 50% decelerating, Sine, 100% decelerating

Standards and Approvals

Classification

Class IIb, continuously operating according to EC directive 93/42/EEC

Certification

IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2014, ANSI/AAMI ES60601-1:2005/(R)2012, ISO 80601-2-12:2011, CAN/CSA-C22.2 NO. 60601-1:14, EN ISO 5356-1:2015, ISO 80601-2-55:2011

Declaration

The HAMILTON-C3 was developed in accordance with pertinent international standards and FDA guidelines. The ventilator is manufactured within an EN ISO 13485 and EN ISO 9001, Council Directive 93/42/EEC, Annex II, Article 3 certified quality management system. The ventilator meets the Essential Requirements of Council Directive 93/42/EEC, Annex I

Electromagnetic Compatibility

According to IEC 60601-1-2:2014

Safety Class

Class II, Type B applied part (ventilator breathing system, VBS), type BF applied parts CO2 sensor including CO2 module connector, humidifier, nebulizer, and SpO2 sensor including SpO2 adapter, continuous operation according to IEC 60601-1

Pneumatic Performance

Oxygen Pressure

2.8 to 6 bar / 41 to 87 psi

Oxygen Connector

DISS (CGA 1240) or NIST

Air Supply

Integrated ultra-quiet turbine

Inspiratory Outlet (To Patient Port)

Connector: ISO 15 mm ID/22 mm OD conical

Expiratory Outlet (From Patient Port)

Connector (on expiratory valve): ISO 15 mm ID/22 mm OD conical

Graphical Patient Data

Intelligent Panel

Dynamic Lung2, Vent Status, ASV Graph3

Waveform

Pressure, Flow, Volume, PCO21, FCO21, Plethysmogram1, Ptrachea

Trend

1, 6, 12, 24, or 72-h trend data for a selected parameter or combination of parameters

Loops

Pressure/Volume, Pressure/Flow, Volume/Flow, Volume/PCO21, Volume/FCO21

Additional Details

Power Supply

100 to 240 V AC, 50/60 Hz or 12 to 24 V DC

Display Size

12.1 inch (307.3 mm) diagonal

Display Type

Color TFT

Battery Backup

3.5 hours with one battery / ≥7 hours with two batteries

Weight

Ventilation unit: 9.5 kg (21 lb); with trolley: 37 kg (81.6 lb)

Alarms

High: Apnea time (s), Exp ManVel high/low (l/min), Oxygen high/low (%), Pressure high/low (cmH2O), Flow sensor calibration needed, Exhalation obstructed, Disconnection, Oxygen supply failed, Medium: Total high/low (b/min), PetCO2 high/low (mmHg), Pressure limitation (cmH2O), Vt high/low (ml), SpO2 high/low, SpOC high/low,%leak, High PEEP, Loss of PEEP, Pulse high/low, Check flow sensor for water, Low: High SpO2, Loss of external power, Cuff leak

I:E Ratio

Adult/Ped: 1:9 to 4:1, Neonatal: 1:9 to 4:1

Automatic Expiratory Base Flow

Fixed at 6 L/min

Means of Inspiratory Triggering

Flow trigger or pressure trigger control

Means of Expiratory Triggering

Flow cycle (ETS)

Minimum expiratory time

20% of cycle time; 0.2 to 0.8 s

Oxygen Mixer Accuracy

± (Volume fraction of 2.5% + 2.5% of actual reading)

Preoperational Checks

Tightness test, Flow sensor/O2 sensor/CO2 sensor calibration

Features

HIgh Flow Oxygen Therapy

High Performance Non Invasive Ventilation

P/V Tool Pro for lung assessment and recruitment

Transpulmonary pressure measurement

Integrated High Performance Turbine

Automated ventilation with INTELLiVENT-ASV

Applications

ICU, Critical Care and Transport

Product Description

The HAMILTON-C3 includes Hamilton Medical's standardized Ventilation Cockpit user interface and the unique intelligent ventilation mode, Adaptive Support Ventilation (ASV). The HAMILTON-C3S edition includes the optional INTELLiVENT-ASV mode. This unique mode takes the proven concepts of ASV one step further: INTELLiVENT-ASV automatically controls the patient's ventilation and oxygenation based on targets set by the clinician and on physiologic input from the patient. The HAMILTON-C3 includes the optional Protective Ventilation Tool (P/V Tool Pro) to assess recruitability. The P/V Tool Pro also provides an easy and repeatable method for performing recruitment maneuvers. The HAMILTON-C3 optionally provides an integrated high flow oxygen therapy mode with a flow of up to 80 l/min.

Write a review

Note: HTML is not translated!
    Bad           Good

C3 VENTILATOR HAMILTON

  • Brand: HAMILTON
  • Product Code: C3
  • Availability: In Stock
  • ₹0Rs.