• C1 NEO  VENTILATOR HAMILTON


Product Details
Brand
Hamilton
Model Name/Number
C1 Neo
Patient Age Group
Neonatal
Tidal Volume
2 to 300 ml
Ventilation Mode
PCV+, PSIMV+, SPONT, APRV, DuoPAP, (S)CMV+/APVcmv, SIMV+/APVsimv, NIV, NIV-ST, nCPAP, nCPAP-PC, HiFlowO2
Optional Enhancements
Monitoring
Display of more than 50 monitoring parameters
Parameter
PEEP/CPAP
3 to 25 cm H2O
Alarm
Loudness
Adjustable (1 - 10), configurable minimum loudness
Environmental
Altitude
Upto approx. 4000 m (13120 feet) 600 to 1100 hPa
Event Log
Storage and display of up to 1000 events with date and time stamp
IntelliTrig
Automatic response to varying leaks and adaption of trigger sensitivity in all modes
PSync
Guaranteed rate ventilation
Standards and Approvals
Certification
IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2007, ISO 80601-2-12:2011 + Cor.:2011, CAN/CSA-C22.2 No. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012
Degree of Protection
IP21
Pneumatic Performance
Air Supply
Integrated turbine
Oxygen Supply
280 to 600 kPa (41 to 87 psi), V max 200 L/min
Low Pressure Oxygen
≤15 L/min, max. 600 kPa (87 psi) for low pressure
Peak Flow
Upto 40 L/min
Graphical Patient Data
Waveform
Paw, Flow, Volume, Plethysmogram, Capnograph
Ventilation Cockpit
Vent Status
Visual representation of ventilator dependency, grouped into oxygenation, CO2 elimination, patient activity
Additional Details
Operating Temperature
5ºC to 40ºC (41ºF to 104ºF)
Power Supply
Input Voltage - 100 to 240 V AC to -15%/+10%, 50/60 Hz, Power Consumption - 50 VA typical, 150 VA maximum
Display Size
8.4 inch
Display Type
Color TFT, backlit touch screen
Battery Backup
Typical 3 hours 45 min, maximum 4 hours 15 minutes
Weight
4.9 kg (10.8 lb) without trolley
Alarms
Low/high minute volume, low/high pressure, low/high tidal volume, low/high rate, apnea time, low/high oxygen, low/high PetCO2, low/high SpO2, low/high pulse, low/high perfusion index, high flow, low/high PVI, low/high SpCO, low/high SpMet, low/high SpHb; Special Alarms: O2 cell, disconnection, exhalation obstructed, loss of PEEP, pressure not released, flow sensor, expiratory valve, pressure limitation, performance limited, CO2 and SpO2, Battery, Power Supply, gas supply, oxygen concentration, check patient interface
Dimensions
136 x 63 cm (53.5 x 24.8 inch)
Relative Humidity
10 - 95% non condensing (operating and storage)
Storage Temperature/ Humidity
-20ºC to 60ºC (-4ºF to 140ºF)
Optional
Humidifier support, cylinder holder, tubing support arm, Volumetric mainstream capnography, sidestream capnography, DuoPAP/APRV, NIV/NIV-ST, Trends/Loops, nCPAP/nCPAP-PC; SpO2, HiFlowO2
Communication/Interface
USB, COM1 (RS-232), Nurse call, CO2, SpO2
Oxygen Inlet
DISS or NIST male
I:E Ratio
1:9 to 4:1 (DuoPAP 1:599 to 149:1)
Main Patient Outlet
ISO 5356-1; 22 OD / 15 ID (mm)
Low Oxygen Inlet
CPC quick coupling, 3.2 mm ID
Features
HIgh Flow Oxygen Therapy
High Performance Non Invasive Ventilation
Mobility for intra-hospital transport
State-of-the-art invasive ventilation modes
Integrated High Performance Turbine
Leak compensation in every mode
Applications
ICU, Critical Care and Transport
Product Description
The HAMILTON-C1 neo is a versatile neonatal ventilator that combines invasive and noninvasive modes with the additional options of nCPAP and high flow oxygen therapy. The integrated turbine allows it to be operated independently of a compressed air supply. Due to its compact design, it is an ideal companion for your smallest patients in various environments such as delivery room, the intensive care unit and emergency ward, as well as during intrahospital transport.